My topload readings are low. Print this pagePrint this page

My topload readings are low.

First try a dead weight test to check calibration. If values are correct check setup menu 3. Select RESULTS setup. Both force and deflection displays should be set to peak. If the force display is set to break during the test the force may increase then decrease before the test stops. At the end of the test the force is displayed at the break instead of the peak.